2 Technical Data
Measurement range
"Validity"
Accuracy
1)
0.20% of reading
Temperature effect on zero
10 Torr/mbar (F.S.)
1 Torr/mbar (F.S.)
0.0050% F.S./ °C
0.015% F.S./ °C
Temperature effect on span 0.01% of reading / °C
Resolution
0.003% F.S.
Gas type dependence none
Output signal analog
(measuring signal)
Voltage range –5 … +10.24 V
Measuring range 0 … +10 V
Relationship voltage-pressure linear
Output impedance
0 Ω (short-circuit proof)
Loaded impedance
>10 kΩ
Response time
30 ms
Gauge identification Resistance 13.2 kΩ refer-
enced to supply common
(voltage at pin 10 5 V)
1)
Non-linearity, hysteresis, repeatability in the calibrated range at 25 °C
ambient operating temperature without temperature effects after
operation of 2 h.